| Display |
| Screen Type:High-resolution display, TFT colour display |
| Screen Size: 7 inches (17.7cm) diagonally |
| Sweep Speed: 12.5/25/50mm/sec |
| Information: Heart Rate, Lead, Alarm On/Off, Delivered
Energy |
| Resolution:800X480 |
| ECG |
| Patient Connection: 3-lead ECG cable, or 5-lead ECG cable,
paddles. |
| Lead Selection: Displayed on monitor, paddles, I, II,
III,AVR, AVL, AVF, V. |
| ECG Size: 2.5,5, 10, 20, 30 mm/mV display on monitor. |
| Heart Rate: 20-350BPM. |
| Protection: With stand 4000vAc/50Hz voltage in isolation and
work against electrosurgical knife and defibrillation |
| Defibrillator |
| Waveform: Biphasic |
| Defibrillator type: Manual External Synchronous/Asynchronous
Defibrillation. |
| Charge Time: Less than 10 seconds to 360 Joules |
| Energy Display: Monitor display indicates both selected and
delivered energy. |
| Charge Prompt Type: Voice and visual prompts. |
| Energy Selection: Selectable at 0, 3, 5, 7, 10, 20, 30, 50,
100, 200, 300, 360 joules |
| Standard Adult/Pediatric Paddles: Reusable external adult
paddles (pediatric paddles integrated) |
| Battery Description |
| Standard Type: Li-ion battery Li1104c 11.1Vdc 4000mAh X2 |
| Standard Capacity: Up to 10 hours ECG monitoring or 60
full-energy discharges. |
| Low Battery Indicator: Flashing low battery icon on display
and Bat Indicator on the panel of defi-monitor be on |
| Recharge Rate: 80% in 3 hours. 100% in 5 hours. |
| Built-in Recorder |
| Paper: 50mm thermal. |
| Printing Method: High-resolution, thermal print head. |
| Record date, time ,sweep speed, waveform gain, heart rate,
energy (joule) and ECG waveform |
| Print mode |
| HR exceed limits alarm recording |
| Auto timed recording |
| Real-Time Recording |
| Defibrillation discharge trigger recording |
| Power |
| AC-DC dual-use: Use the batteries,AC100V~240V,50/60 Hz. |
| Vehicle Voltage(Option): DC 12V |
| Main unit: |
| Portable |
| Maximum power 600W |
| Dimensions:L320xW205xH410 |
| Weight:7.5kg |
| Pulse Oximeter (Optional) |
| Display: Spo2 value,waveform, Pulse |
| Probe: Standard adult finger clip |
| Range: 0-99% for adult, pediatric and neonatal patients |
| AED (Optional) |
| Voice prompts and screen prompts. |
| Energy Programmable Sequence: |
| 200,300,360 Joules; |
| 150,150,200 Joules; |
| 50, 75, 100 Joules; |
| 30, 30, 50 Joules; |
| Discharge Control: Control on device front panel. |